 1、Background

Electric shock and system fault caused by power distribution system are mostly caused by the grounding fault of equipment line. Grounding fault will directly bring the harm of power supply interruption; when the earth fault leakage is serious, it will even burn electrical equipment or lines, causing irreparable loss.

 

In Medical places, especially those life-threatening class 2 medical places, patients in surgery or anesthesia state, without self-behavior ability, small leakage current can lead to electric shock death of patients. Therefore, the electrical design of this special place should be carried out strictly in accordance with the relevant national standards and specifications.

IT power distribution system (also known as isolated power supply system or ungrounded system) has very small leakage current to ground due to its neutral point ungrounded. When the first insulation fault occurs, the power supply of the equipment will not be affected. So the system can continue to operate. The insulation monitoring device installed in the system can detect the fault in time and send out the corresponding fault warning, prompting the relevant personnel to remove the fault at the appropriate time. For the operation room, intensive care unit and other medical places with high requirements for power supply continuity and safety, local IT system power supply can ensure the security and reliability of power supply for important loads.

 

2、Project Information

Ky Anh's Medical Center, a medical facility in Vietnam, has introduced five sets of insulation monitoring products for operating rooms to improve its medical facilities.

3、Description of four sets of medical IT products

The medical IT system is mainly used in important Class 2 medical locations such as operating room, ICU/CCU intensive care unit, providing safe, reliable and continuous power distribution for the important equipment at these locations. Medical insulation monitoring and fault locating device is developed by the many years’ design experience of the Acrel Electric in electric power meter industry, according to the special requirements of the insulation monitoring and fault locating of the power distribution system in Class 2 medical locations. The products can realize the real-time monitoring of insulation, load and temperature of isolation transformer in IT system, and have the functions of system insulation fault loop positioning and centralized monitoring by multiple pieces of systems. Products conform to the provisions of enterprise standard Q31/0114000129C013-2016 IT System Insulation Monitoring Instrument.

Medical IT system insulation monitoring (four-piece set) include AITR series medical isolation transformer, AIM-M10 medical intelligent insulation monitoring instrument, AKH-0.66P26 current transformer, and AID series(AID10、AID130、AID150) centralized alarm and display instrument, which are shown in Table 1 

4、Typical application of isolated power supply system

作者简介:

任华,女,本科 现任于安科瑞电气股份有限公司

联系电话:021-59104851 

 

手机:187-0211-2061 QQ: 2880157851

北极星光伏商务通微信

北极星电力商务通微博

扫一扫
关注北极星商务通官方微信公众号及微博,及时获取最新资讯。

北极星电力网声明:此资讯系转载自北极星电力网合作媒体或互联网其它网站,北极星电力网 登载此文出于传递更多信息之目的,并不意味着赞同其观点或证实其描述。文章内容仅供参考。

供应商档案

安科瑞电气股份有限公司
星级
企业认证: 企业已认证已认证
所在地区: 上海
主营产品:

环保用电监测云平台、智慧安全用电云平台、智慧消防云平台、水电预付费云平台、电量传感器、隔离电源柜、光伏汇流箱、低压有源滤波装置、消防电子仪表及软件保护装置、计算机通讯柜的研发、制造、销售,半导体照明的研发,销售,合同能源管理、节能技术检测、电力监控、电能节能管理、建筑能耗检测、电气火灾监控领域内的技术开发、技术转让、技术咨询、技术服务、信息系统集成及软件开发,从事货物和技术的进出口业务。

收藏此企业